Conductometer — concentration
meter two channel with contact and inductive sensors
C-3122.x, C-3122.x.I-Ех
The analyzer is a two-channel measuring device which consists of
one or two remote active primary transducers1)
(PT) of the specified electrical conductivity (SEC) and a two-channel measuring instrument (MI).
The length of the three-wire communication line between the PT and the MI can reach 1000 m.
The analyzer is designed to measure SEC, temperature
and concentration of salts, alkalis and acids solutions.
The analyzer PT can be equipped with contact or
inductive proximity sensors.
The analyzer provides a digital display of SEC values and
temperature, their conversion into proportional values of uniform DC output signals, data exchange via digital
interface RS-485, alarming when the measured parameters are out of the specified values, as well as the
measurement results archiving and graphical display.
C-3122.х.I-Ех analyzers (PT in the “I”enclosure)
with a marking 1Ex d IIB T6 X have an explosion protection type "flameproof enclosure". Accessories:
The flow rate of the liquid for the flowing contact sensor
max 100 l/h
Linear fluid velocity for submersible sensor
max 0,5 m/sec
Viscosity of the analyzed liquid
max 0,2 Pa*sec
Output signal from PT
digital impulse current
Degree of protection against water and dust:
IP65
Resistanse to electromagnetic influence:
IV by GOST 32137 (Rus), criterion A
Climatic version:
(-40..+50)°C
Resistance to mechanical influences in accordance with GOST R 52931 (Rus)
V2
Weight with flowing sensor
max 2,5 kg
MEASURING INSTRUMENT
Quantity of measuring channels
2
Parameters being measured in every channel
SEC and temperature
Communication line length from the PT to MI
max 800 m
Measuring range (according to analogue output signal)
set programmatically
Indicator type
LCD graphics
Output signals:
- direct current analogue, proportional to the measuring ranges of SEC and temperature
galvanically isolated from the input signal
(0...5), (0...20) mА or (4...20) mА (selected programmatically)
- digital interface
RS-485, Modbus RTU data communications protocol
- discrete, programmable, actuation according to SEC or temperature set points
four relay with switching contacts, ~240V, 3A
Archiving time
from 4.4 hours to 55 days
Archiving interval
programmable from 1 sec to 5 min
Power supply
~(100...240) V, (50...60) Hz
Power consumption
max 15 VA
MI enclosure material:
- panel mounted design
aluminum alloy
- wall mounted design
ABS plastic
Climatic version:
T=(+5..+50)°C
Dust and water protection:
- wall mount MI enclosures
IP65
- panel mount MI enclosures on the front panel
IP54
Resistance to mechanical influences in accordance with GOST R 52931 (Rus)
N2
Weight:
max 1 kg
1)In C-3122 it is possible to programmatically switch channels to work with the primary converters of the device pH-4121. 2)Upper measuring limit for contact submercible sensors 100 mS/cm.. 3)By special order, the contact sensor for C-3122.1 is manufactured for temperature of up to 120 °С (HT). The upper limit of
the temperature of the liquid to be analyzed is determined depending on the medium. 4)The reference temperature for termocompencation (°C) and the temperature coefficient (% per °C) are set